6 PCD-BA-e-12112 Basic Information (General)2.1 Proper useThe balance you purchased is intended to determine the weighing value of materialto be weighed. It is intended to be used as a “non-automatic balance”, i.e. thematerial to be weighed is manually and carefully placed in the centre of the weighingplate.. As soon as a stable weighing value is reached the weighing value can beread.2.2 Improper UseDo not use balance for dynamic weighing. In the event that small quantities areremoved or added to the material to be weighed, incorrect weighing results can bedisplayed due to the “stability compensation“. (Example: Slowly draining fluids from acontainer on the balance).Do not leave permanent load on the weighing plate. This may damage themeasuring system.Impacts and overloading exceeding the stated maximum load (max) of the balance,minus a possibly existing tare load, must be strictly avoided. Balance may bedamage by this.Never operate balance in explosive environment. The serial version is not explosionprotected.The structure of the balance may not be modified. This may lead to incorrectweighing results, safety-related faults and destruction of the balance.The balance may only be used according to the described conditions. Other areas ofuse must be released by KERN in writing.2.3 WarrantyWarranty claims shall be voided in caseOur conditions in the operation manual are ignoredThe appliance is used outside the described usesThe appliance is modified or openedMechanical damage and damage caused by media, liquidsnatural wear and tearThe appliance is improperly set up or incorrectly electrically connectedThe measuring system is overloaded2.4 Monitoring of Test ResourcesIn the framework of quality assurance the measuring-related properties of thebalance and, if applicable, the testing weight, must be checked regularly. Theresponsible user must define a suitable interval as well as type and scope of thistest. Information is available on KERN’s home page (www.kern-sohn.com) withregard to the monitoring of balance test substances and the test weights required forthis. In KERN’s accredited DKD calibration laboratory test weights and balances maybe calibrated (return to the national standard) fast and at moderate cost.