46 PFB-BA-e-132210.4 Continuous data output, -/U k g CR LF-Header 1- -Header 2- -- Weighing data -- Weighing unit Final characterHead line 1: ST=stable, US=instableHead line 2: NT=net, GS=gross10.5 Remote control instructionsThe remote control commands are sent from the remote control unit to the balanceas ASCII code. After the balance having received the s/w/t commands, it will send thefollowing data.T Function: Tare balanceZ Function: Set balance to zeroW Function: Weighing value for the weight (stable or unstable) is sentvia the RS232 interfaceS Function: Stable weighing value for the weight is sent via the RS232interfaceP Function: In counting mode the number of the pieces to be counted11 Error messagesErr 3 Incorrect adjusting weightPut on correct adjustmentweight (see chap. 1; TechnicalDataErr 4 Zero range exceeded Remove load and press toreset balance to zero.Err 5 Keyboard error Inadmissible inputErr 6Electronic errorSwitch balance off and onagain. If the error messageremains displayed, pleasecontact your dealer.Transportation lock Remove the transportation lock