48 PFB-BA-e-132213 Instant helpIn case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Help:Fault Possible causeThe displayed weight doesnot glow. The balance is not switched on.The mains supply connection has been interrupted(mains cable not plugged in/faulty).Power supply interrupted.(Rechargeable) batteries are inserted incorrectly oremptyNo (rechargeable) batteries inserted.The displayed weight ispermanently changing Draught/air movementTable/floor vibrationsWeighing plate has contact with other objects.Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)The weighing result isobviously incorrect The display of the balance is not at zeroAdjustment is no longer correct.Great fluctuations in temperature.Warm-up time was ignored.Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.