8.2 General functionsMenu selection:Ö Turn on balance using the ON/OFF-buttonÖ Press the Tare-button; the display shows horizontal linesÖ While this display is on, press the Cal-buttonÖ Keep F-button pressed until the desired selection is displayedFunction DisplayF-buttonSelectionPRINT-buttonF-buttonDescription of theselection options0 offRear illuminatedDisplay(see Chapter. 7.1.1)bl1 onAUTO 0* Auto-zero-function activatedAuto-zero-function(see Chapter 6.5.4) AUTOAUTO 1 Auto-zero-function deactivatedAc_t 0* Automatic adjustment functionactivatedAc_tAc_t 1 Automatic adjustment functiondeactivatedAC_c 0* Automatic adjustment functionactivatedSetting the automaticadjustment function,only ALJ /PLJ6100-2in non-verified setting(see Chapter 6.7.1)Ac_cAC_c 1 Automatic adjustment functiondeactivatedPd_d 0* Last digit is not printedData output(only ALJ / PLJ6100-2in non-verified setting)Pd_dPd_d 1 Last digit is printedVibration filterConCon 1Con 2Con 3*Con 4Con 5°Sensitive and fast(very quiet set-up location)Robust but slow (very busyset-up location)Forming of anaverage value(Animal weighingprogram)AuEAuE 1AuE 2*AuE 3Fastest displaySlowest displayFIL 0* Setting for dosingdeactivatedDisplay speedFILFIL 1 Setting for dosingactivatedStandard weighingunit St_uni see Chapter 7.2.2* = Factory setting° = Recommended counter settingPLS/PLJ/ALS/ALJ-BA-e-081962