PLT-BA-defsi-051088Ö With "Print report" activated, you will receive a printout of your adjustment dataafter each adjustment (example):8.3.1.1 VerificationGeneral introduction:According to EU directive 90/384/EEC balances must be officially verified if they areused as follows (legally controlled area):a) For commercial transactions if the price of goods is determined by weighingb) For the production of medicines in pharmacies as well as for analyses in themedical and pharmaceutical laboratoryc) For official purposes.d) For manufacturing final packages.In cases of doubt, please contact your local trade in standard.Verification instructionsAn EU type approval exists for balances described in their technical data as verify-able. If a balance is used where obligation to verify exists as described above, itmust officially verified and re-verified in regular intervals.Re-verification of a balance is carried out according to the respective national regula-tions. The validity for verification of balances in Germany is e.g. 2 years.The legal regulation of the country where the balance is used must be observed!After verification the balance is sealed at the indicated positions.Verification of the balance is invalid without the "seal".***Print report of internal calibration***Calibrate : internalDiff.. : 0.0038 gSignature………………………….