Instructions for use ELECTROmatic M/C and PM/PC4 Installation | 4.1 Location32 / 804 Installation4.1 LocationCAUTIONDamage to the dentist element.Installations involving an intervention on the dental unit might damage compo-nents, which can interfere with the safe function and cause injury.▶ Have installations involving an intervention on the dental unit performed bytrained expert personnel only.▶ Have the treatment center subjected to a safety check after installation.CAUTIONPower supply and cords/hoses are on the floor.Slipping and tripping.▶ Power supply and cord/hoses need to be set-up/routed appropriately suchthat they are not on the floor.NoteTo completely disconnect the device from the mains, the mains plug must bepulled. For this reason, the unit must be set-up appropriately such that themains plug and the electrical outlet are easily accessible.NoteMind the electrical cord of the power supply! Route all cords appropriatelysuch that they do not get squashed, clamped or have a chair rolling overthem.NoteUse the power supply in dry environments exclusively. Make sure that thepower supply is protected from the ingress of liquids.▶ Place the product in an easily accessible place and visible for diagnostic pur-poses.4.2 Installation positionsThere is a wide range of installation options for the ELECTROmatic. This over-view makes no claim of being comprehensive. If possible, the installationshould be done without modification of the dentist element and/or existing at-tachment options on the dentist element should be utilized.The 4 basic installation positions on holders are described in the following:▪ Installation position 1: Below a holder▪ Installation position 2: On the side of a holder▪ Installation position 3: On a holder or on the backside of a holder▪ Installation position 4: Mount control panel as remote control