Instructions for use ELECTROmatic M/C and PM/PC12 Information about electromagnetic compatibility | 12.3 Guidelines and manufacturer's statement - Electromag-netic immunityInterference im-munity testsIEC 60601 testlevelsCompliance level Electromagnetic environment- GuidelinesElectrostatic discharge(ESD) according toIEC 61000-4-2± 8 kV contact dis-charge± 15 kV atmosphericdischarge± 8 kV contact dis-charge± 15 kV atmosphericdischargeFloors should be made of woodor concrete or be fitted withceramic tiles. If the floor is fittedwith synthetic material, the relat-ive humidity must be at least 30%.Fast transient elec-trical interference /bursts according toIEC 61000-4-4± 2 kV for power lines± 2 kV for input andoutput lines± 2 kV for power linesand input and outputlinesThe quality of the supply voltageshould correspond to that of atypical business or hospital envir-onment.Surges according toIEC 61000-4-5± 1 kV push-pullvoltage± 2 kV common modevoltage± 1 kV push-pullvoltage± 2 kV common modevoltageThe quality of the supply voltageshould correspond to that of atypical business or hospital envir-onment.Voltage interruptions,short-term interrup-tions, and fluctuationsof the supply voltageaccording to IEC61000-4-110 % / 0.5 cyclesat 0º to 315º in 45ºincrements0 % / 1 cycle70 % / 25 cycles0 % / 250 cycles0 % / 0.5 cyclesat 0º to 315º in 45ºincrements0 % / 1 cycle70 % / 25 cycles0 % / 250 cyclesThe quality of the supply voltageshould correspond to that of atypical business or hospital envir-onment.Magnetic field at asupply frequency(50/60 Hz) accordingto IEC 61000-4-830 A/m 30 A/m Magnetic fields at the mains fre-quency should correspond to thetypical values in a business andhospital environment.12.3 Guidelines and manufacturer's statement -Electromagnetic immunityThe ELECTROmatic is intended for use in an environment of the kind specifiedbelow. The customer or user of the ELECTROmatic should make sure that thedevice is used in an environment of the specified type.73 / 78