2 Introduction• All electronic circuits and electronic boards• Plastic parts2.8 Warnings and requirements2.8.1 Warnings and precautions for use• Before using the device for the first time, familiarize yourself with this manual toensure the safe use of the device.• Before using the device for the first time, ensure that it has been set up according toyour requirements.• This device complies with the EMC (Electromagnetic Compatibility) according to IEC60601-1-2. Radio transmitting equipment, cellular phones etc. shall not be used inproximity of the device as they could influence the performance of the device.• Do not move the upper shelf of the device by hand as it may hit a wall or cause thedevice to go to an error state.• If the patient is using a pacemaker, consult the manufacturer of the pacemaker beforetaking an X-ray image to confirm that the device will not interfere with the operationof the pacemaker.• Always ensure the patient positioning before taking X-ray images to diminish the needfor retakes.• When taking an X-ray image, protect yourself from radiation or stand at least 3meters (9.8 ft) away from the device.• When taking an X-ray image of a patient who is feeling insecure or has exceptionalanatomy (typically very tall or large patient) use the Test-mode first to demonstratethe device movements and to make sure that the rotating device doesn't collide withthe patient during the imaging process.• Avoid taking X-ray images of pregnant women.• Always ensure that you can see and hear the patient during the imaging process,while also maintaining visibility to the GUI.• Always place new disposable covers on patient positioning parts and decontaminatethe patient contacting parts according to guidance in chapter Cleaning anddecontamination.• Close the patient positioning mirror before taking images.• Switch the device off after use.• The device must not be used for screening examinations.2.8.2 Warnings and precautions for installation• Only authorized service technicians are allowed to perform the device installation orservice.• The column telescope assembly is not a serviceable part and it should not be openedor altered in any circumstances.• Always ensure that the device is isolated from mains during any installation or serviceprocedures to avoid a risk of electric shock. Live parts are marked with high voltagesymbols.• Familiarize yourself with this manual before device installation and follow theinstallation instructions to ensure the safe use of the device.ORTHOPANTOMOGRAPH™ OP 3D 9