Instructions for use SMARTmatic PROPHY S31 - 1.011.6760 SMARTmatic PROPHY S31 K - 1.011.6761 SMARTmaticPROPHY S33 - 1.011.68007 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection18 / 267.3.1 Manual external disinfectionKaVo recommends the following products based on compatibility of the materi-als. The microbiological efficacy must be ensured by the disinfectant manufac-turer and proven by an expert opinion.▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)▪ FD 322 made by Dürr▪ CaviCide made by MetrexConsumables required:Cloths for wiping the medical device.▶ Spray the disinfectant on a cloth, then wipe down the medical device andallow the disinfectant to act according to the instructions of thedisinfectant manufacturer.▶ Follow the instructions for use of the disinfectant.7.3.2 Manual internal disinfectionThe efficacy of manual internal disinfection must be demonstrated by the man-ufacturer of the disinfection agent. With KaVo products, use only disinfectionagents that have been released by KaVo with respect to the compatibility ofmaterials (e.g. WL-cid / made by ALPRO).▶ Immediately after internal disinfection, lubricate the KaVo medical devicewith care agents from the KaVo care system.7.3.3 Automated external and internal disinfectionKaVo recommends washer disinfectors from Miele in accordance with EN ISO15883-1 that are operated with alkaline cleaning agents.The validations were conducted with the VARIO-TD programme, the cleaningagent neodisher® MediClean and the neutralisation agent neodisher® Z.▶ For programme settings as well as cleansers and disinfectants to be used,please refer to the Instructions for Use of the washer disinfector.▶ In order to prevent damage to the medical device due to residual fluid,make sure that the inside and outside of the device is dry after the end ofthe cycle. Remove any residual liquids from the interior and exterior of themedical device using compressed air.▶ Immediately after drying, lubricate the medical device with care agentsfrom the KaVo care system.The drying procedure is normally part of the cleaning program of the thermod-isinfector.NotePlease comply with the instructions for use of the washer disinfector.