Instructions for use SMARTmatic PROPHY S31 - 1.011.6760 SMARTmatic PROPHY S31 K - 1.011.6761 SMARTmaticPROPHY S33 - 1.011.68007 Reprocessing steps in accordance with ISO 17664 | 7.6 Packaging20 / 26See also:2 Instructions for use KaVo SPRAYrotor7.5.3 Servicing with KaVo QUATTROcareNoteQUATTROcare 2104 / 2104 A is no longer included in the current de-livery programme.Follow-up product:▶ QUATTROcare PLUS 2124 AServicing and cleaning device with expansion pressure for the interior cleaningof inorganic residues and optimum care.(no validated cleaning of the interior according to German Robert Koch Institute(RKI) requirements)KaVo recommends servicing the product as part of the reprocessing after eachuse, i.e. after each cleaning, disinfection, and before each sterilisation.▶ Take screw-in and/or snap-on polishing attachments off the contra-anglehandpiece.▶ Service the product.See also:2 Instructions for use KaVo QUATTROcare 2104 / 2104A7.5.4 Servicing with KaVo QUATTROcare PLUSServicing and cleaning device with expansion pressure for the interior cleaningof inorganic residues and optimum care.(no validated cleaning of the interior according to German Robert Koch Institute(RKI) requirements)KaVo recommends servicing the product as part of the reprocessing after eachuse, i.e. after each cleaning, disinfection, and before each sterilisation.▶ Take screw-in and/or snap-on polishing attachments off the contra-anglehandpiece.▶ Service the product in the QUATTROcare PLUS.See also:2 Instructions for use KaVo QUATTROcare PLUS7.6 PackagingNoteThe sterilisation bag must be large enough for the instrument so that the bagis not stretched.The quality and use of the sterilisation packaging must comply with applic-able standards and be suitable for the sterilisation procedure!▶ The medical device must be packed before sterilisation.