Keysight 16860 Series Portable Logic Analyzer Service Guide 35Testing 16860 Performance 3Test the 16860 series Logic AnalyzerThe following sections explain how to measure the maximum state data rate.1 Record the logic analyzer's model and serial number in the Performance Test Record (seepage 77). Record your work order number (if applicable) and today's date.2 Record the test equipment information in the “Test Equipment Used” section of the PerformanceTest Record.3 Turn on the frame.a Plug in the power cord to the power connector on the rear panel of the logic analyzer.b Connect a key board and a mouse to the frame.c Press the Standby (Power) button on the front panel of the logic analyzer to power on the logicanalyzer.While the logic analyzer is booting, observe for the following:• Ensure all of the installed memory is recognized.• Any error messages.• Interrupt of the boot process with or without error message.4 During initialization, check for any failures.If an error or an interrupt occurs, refer to the chapter “Troubleshooting" on page 81.5 Start the Keysight Logic Analyzer application if it is not started already.NOTE Before testing the performance of the logic analyzer, warm-up the logic analyzer and the testequipment for 30 minutes.