85052B 2-5SpecificationsElectrical SpecificationsElectrical SpecificationsThe electrical specifications in Table 2-3 apply to the devices in your calibration kit when connected with aKeysight precision interface.Table 2-3 Electrical Specifications for 85052B 3.5 mm DevicesDevice Specification Frequency (GHz)Broadband loads Return loss 46dB 0.00501) dc to 2(male and female) Return loss 44 dB 0.00631) > 2 to 3Return loss 38dB 0.01259) > 3 to 8Return loss 36dB 0.01585) > 8 to 20Return loss 34dB 0.01995) > 20 to 26.5Sliding loads aa. Assuming proper usage, the specifications for the residual return loss after calibration for the slidingload termination include:• the quality of the airline portions within the sliding load, combined with • the effective stability of the sliding element.Proper usage includes the following practices: • Connector mating surfaces are clean.• The changes in slide positioning are NOT done in equal steps since this results in very poorcalibration for some portions of the frequency range.• The center conductor of test port connectors are nominally set back from the outer conductor. Sliding loads are designed to allow the center conductor to be moved. The position of thesliding load center conductor should be set by a reference block and not positioned flush againstthe center conductor of the test port.Return loss 44 dB 0.00631) > 3 to 26.5Offset opens b 0.65 deviation from nominal dc to 3(male and female) 1.20 deviation from nominal > 3 to 8 2.00 deviation from nominal > 8 to 20 2.00 deviation from nominal > 20 to 26.5Offset shorts b 0.50 deviation from nominal dc to 3(male and female) 1.00 deviation from nominal > 3 to 8 1.75 deviation from nominal > 8 to 20 1.75 deviation from nominal > 20 to 26.5Adapters c Return loss 30 dB 0.03162) dc to 8Return loss 28 dB 0.03981) > 8 to 18Return loss 26 dB 0.05012) > 18 to 26.5