46 6000L Series Oscilloscopes Service Guide2 Testing PerformanceKeysight 6000L Series Oscilloscopes Performance Test RecordSerial No. ______________________________________ Test by _____________________________Test Interval ____________________________________ Work Order No. ______________________Recommended Next Testing ________________________ Temperature ____________Threshold Specification Limits Ch D7-D0 Ch D15-D8Accuracy Test 5 V - 250 mV 4.750 V ________ ________(100 mV + 3% ofthreshold setting)5 V + 250 mV 5.250 V ________ ________-5 V - 250 mV -5.250 V ________ ________-5 V + 250 mV -4.750 V ________ ________0 V - 100 mV -100 mV ________ ________0 V + 100 mV 100 mV ________ ________Voltage Measurement AccuracyRange Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3* Channel 4*5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________2 mV/Div (500 MHz and 1 GHz models)14 mV 13.232 mV to 14.768 mV ________ ________ ________ ________2 mV/Div (100 MHz models)14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________1 mV/Div (MSO/DSO6012A)7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________Bandwidth Model Test Limits Channel 1 Channel 2 Channel 3* Channel 4*610x -3 dB at 1 GHz ________ ________ ________ ________605x -3 dB at 500 MHz ________ ________ ________ ________601x -3 dB at 100 MHz ________ ________ ________ ________Time ScaleAccuracyLimitsTime Scale Accuracy is a warranted specificationfor 6000L Series instruments manufactured afterJanuary 1, 2008.CalculatedTSA Limit(ppm)Measuredtime scaleerror (ppm)Pass/FailTSA Limit: ± (15 + 2 * YrsSinceMfr) ppm ___________ ___________ ___________