N6954A/N6974A with Option 301N6954A/N6974A Test Record Report Number ______ Date__________Test Description Model Min. Specs Results Max. SpecsVoltage Programming & ReadbackMinimum voltage (Vout):Voltage measured over interface:High voltage (Vout):Voltage measured over interface:BothBothBothBoth+ 72 mVVout – 8 mV79.968 VVout – 32 mV____________________________+ 88 mVVout + 8 mV80.032 VVout + 32 mVCV Load Effect: Both – 2 mV _______ + 2 mVCV Ripple and Noisepeak-to-peak:rms:BothBothN/AN/A______________+ 9 mV+ 1 mVTransient response @ 100 μs: Both – 200 mV _______ + 200 mVCurrent Programming & ReadbackMinimum current (Iout):Current measured over interface:High current (Iout):Current measured over interface:N6954AN6974AN6954AN6974AN6954AN6974AN6954AN6974A– 2 mA– 4 mAIout – 2 mAIout – 4 mA12.493 A24.986 AIout – 0.007 AIout – 0.014 A________________________________________________________+ 2 mA+ 4 mAIout + 2 mAIout + 4 mA12.507 A25.014 AIout + 0.007 AIout + 0.014 ALow Current Measurement Range: N6954AN6974AIout – 0.000875 AIout – 0.0018 A______________Iout + 0.000875 AIout + 0.0018 ACC Load Effect: N6954AN6974A– 0.8 mA– 1.5 mA______________+ 0.8 mA+ 1.5 mACurrent Sink Tests10% of current rating:100% of current rating:N6954AN6974AN6954AN6974A– 1.2525 A– 2.505 A– 12.507 A– 25.014 A____________________________– 1.2475 A– 2.495 A– 12.493 A– 24.986 AN6954ASettings N6974ASettingsVoltage Programming & Readback Min:Voltage Programming & Readback High:CV Load Effect, CV Ripple and Noise:Transient Response:Current Programming & Readback Min:Current Programming & Readback High:Low Current Measurement Range:CC Load Effect:10% of current rating (Isink):100% of current rating (Isink):Voltage priority: 80 mV, 12.5 AVoltage priority: 80 V, 12.5 AVoltage priority: 80 V, 12.5 AVoltage priority: 80 V, 6.25 A to 12.5 ACurrent priority: 80 V, 0 ACurrent priority: 80 V, 12.5 ACurrent priority: 80 V, 1.25 ACurrent priority: 80 V, 12.5 ACurrent priority: 80 V, - 1.25 AExternal supply: 80 V, 1.375 ACurrent priority: 80 V, - 12.5 AExternal supply: 80 V, 13.75 AVoltage priority: 80 mV, 25 AVoltage priority: 80 V, 25 AVoltage priority: 80 V, 25 AVoltage priority: 80 V, 12.5 A to 25 ACurrent priority: 80 V, 0 ACurrent priority: 80 V, 25 ACurrent priority: 80 V, 2.5 ACurrent priority: 80 V, 25 ACurrent priority: 80 V, - 2.5 AExternal supply: 80 V, 2.75 ACurrent priority: 80 V, - 25 AExternal supply: 80 V, 27.5 AKeysight N6900/N7900 Series Operating and Service Guide 3317 Verification and Calibration