N7952A/N7972AN7952A/N7972A Test Record Report Number ______ Date__________Test Description Model Min. Specs Results Max. SpecsVoltage Programming & ReadbackMinimum voltage (Vout):Voltage measured over interface:High voltage (Vout):Voltage measured over interface:BothBothBothBoth+ 36 mVVout – 4 mV39.984 VVout – 16 mV____________________________+ 44 mVVout + 4 mV40.016 VVout + 16 mVCV Load Effect: Both – 1.5 mV _______ + 1.5 mVCV Ripple and Noisepeak-to-peak:rms:BothBothN/AN/A______________+ 9 mV+ 1 mVTransient response @ 100 μs: Both – 100 mV _______ + 100 mVCurrent Programming & ReadbackMinimum current (Iout):Current measured over interface:High current (Iout):Current measured over interface:N7952AN7972AN7952AN7972AN7952AN7972AN7952AN7972A– 4 mA– 8 mAIout – 4 mAIout – 8 mA24.986 A49.972 AIout – 0.014 AIout – 0.028 A________________________________________________________+ 4 mA+ 8 mAIout + 4 mAIout + 8 mA25.014 A50.028 AIout + 0.014 AIout + 0.028 ALow Current Measurement Range: N7952AN7972AIout – 0.0019 AIout – 0.0037 A______________Iout + 0.0019 AIout + 0.0037 ACC Load Effect: N7952AN7972A– 1 mA– 1.5 mA______________+ 1 mA+ 1.5 mACurrent Sink Tests10% of current rating:100% of current rating:N7952AN7972AN7952AN7972A– 2.505 A– 5.01 A– 25.014 A– 50.028 A____________________________– 2.495 A– 4.99 A– 24.986 A– 49.972 AN7952A Settings N7972A SettingsVoltage Programming & Readback Min:Voltage Programming & Readback High:CV Load Effect, CV Ripple and Noise:Transient Response:Current Programming & Readback Min:Current Programming & Readback High:Low Current Measurement Range:CC Load Effect:10% of current rating (Isink):100% of current rating (Isink):Voltage priority: 40 mV, 25 AVoltage priority: 40 V, 25 AVoltage priority: 40 V, 25 AVoltage priority: 40 V, 12.5 A to 25 ACurrent priority: 40 V, 0 ACurrent priority: 40 V, 25 ACurrent priority: 40 V, 2.5 ACurrent priority: 40 V, 25 ACurrent priority: 40 V, - 2.5 AExternal supply: 40 V, 2.75 ACurrent priority: 40 V, - 25 AExternal supply: 40 V, 27.5 AVoltage priority: 40 mV, 50 AVoltage priority: 40 V, 50 AVoltage priority: 40 V, 50 AVoltage priority: 40 V, 25 A to 50 ACurrent priority: 40 V, 0 ACurrent priority: 40 V, 50 ACurrent priority: 40 V, 5 ACurrent priority: 40 V, 50 ACurrent priority: 40 V, - 5 AExternal supply: 40 V, 5.5 ACurrent priority: 40 V, - 50 AExternal supply: 40 V, 55 A7 Verification and Calibration334 Keysight N6900/N7900 Series Operating and Service Guide