60 SMART Spectro 2 Operator’s Manual 04.165. Press or to scroll to the user test tobe edited.EDIT USER TEST 535NM101 USER TEST 1102 USER TEST 2103 USER TEST 3104 USER TEST 4105 USER TEST 5106 USER TEST 6107 USER TEST 7108 USER TEST 86. Press *ENTER to select the user test to beedited.NOTE: The menu allows a test to berenamed, a wavelength to be selected,reacted standards to be scanned and thenumerical format of the displayed testresult to be selected. After editing anyone of these menu choices the displaywill return to this menu. Any menu choicecan be edited at any time by selectingit. The normal procedure would be tostart with NAME THE TEST, then SELECTWAVELENGTH, then NEW CALIBRATION,and then FORMAT RESULT.102 USER TEST 2 535NMNAME THE TESTSELECT WLNEW CALIBRATIONFORMAT RESULTNAMING THE TESTA NAME can be up to 14 characters long. The menu choices for each character are 26letters A to Z, ten numerals 0 to 9, a space, a dash, a decimal point, and a ! as a terminator.Selecting the terminator indicates the end of the name and stores the name. Theterminator, !, is the first menu choice since it is the one character that will always have tobe selected. It is before the letter A. DO NOT CHOOSE THE ! UNTIL THE NAMING OF THE TESTHAS BEEN COMPLETED. The terminator should be selected following the end of the name.It must be the last character selected.info@GlobalScientificSupply.comwww. GlobalScientificSupply .com GLOBAL SCIENTIFIC SUPPLYThe Laboratory Supply Company!