- 25 -8.1.8. Animal and Dynamic Weighing1. To configure the Animal and Dynamic Weighing Function, use the navigation keys to select “AnimalWeighing” in “Mode Setup” and press ENTER. The status and available Animal Weighing optionswill be displayed. If the function is currently in use the status will be set to “ON”.2. To select an operating mode use the navigation keys to select “Mode” and press ENTER. By defaultAnimal Weighing is set to “Auto” (Automatic). To operate Animal Weighing in Semi-Automatic orManual use the navigation keys to select a desired operating mode and press ENTER.Animal Weighing Operating ModesAutomatic Weighing Initialization : AutomaticResult Clearing : AutomaticTare : AutomaticWhen set to Automatic, the scale willautomatically begin the animal weighingprocess. Once the animal is removed fromthe weighing pan, the scale willautomatically tare and prepare for the nextweighing.Semi- AutomaticWeighing Initialization : AutomaticResult Clearing : ManualTare : ManualWhen set to Automatic, the scale willautomatically begin the animal weighingprocess. Once the animal is removed fromthe weighing pan, the result remainsdisplayed until it is manually cleared. Thescale must be manually tared before the nextweighing.ManualWeighing Initialization : ManualResult Clearing : ManualTare : ManualWhen set to Manual, the animal weighingprocess must be manually initialized. Oncethe animal has been weighed and the pancleared the result must be manually cleared.The scale must be manually tared before thenext weighing.