Chapter 4: Lyo-Works Operating SystemProduct Service 816-333-8811 or 1-800-522-7658 27Original instructions6. Maintenance – This screen contains data and settings pertaining to theoperation hours and maintenance period for the vacuum pump and refrigerationsystem. Scroll up and down to view the entire page. Line Voltage Offset - These offset values determine the acceptable voltagerange for electrical power supplied to the Freeze Dryer. If the supply linevoltage is outside of the acceptable range, then the Line Voltage Range alertwill be displayed. The default range is +/- 10% of the nominal rated voltage(115V or 230V depending on model). The LOW Line Voltage Offset can beadjusted from 0 to -10V (115V models) or 0 to -20V (230V models). TheHIGH Line Voltage Offset can be adjusted from 0 to +10V (115V models) or0 to + 20V (230V models). Scroll up and down to view the entire page. Vacuum Pump Protection – If vacuum pump protection is turned ON, thevacuum pump will automatically be turned off if a “Vacuum Leak” alert ispresented. This will protect oil filled pumps from losing their oil andpotentially damaging the pump in the event of a major vacuum leak. Thefactory default setting for vacuum pump protection is ON, and it is highlyrecommended that it remain in the ON setting. If vacuum pump protectionis turned OFF, the pump will continue to run after a “Vacuum Leak” alert ispresented (i.e. there is no protection for the pump in the event of a majorvacuum leak).