ECO GOLD25/08/2011/ YACE0088 Appendix with settings 101− Access to the main menu level is obtained by pressing theenter key .− Selection and confirmation of Æ Setup Æ Control ÆContr.parameter Æ extern Pt100.The adjacent menu window appears. Apart from the control pa-rameters the currently set values are displayed.− Under the menu point "Tv manual/auto" you can select be-tween manual and automatic entry using .The selection "automatic" is displayed in the menu line by (auto).If "automatic" is selected, the entry is blocked for the parametersTv and Td.− Select and confirm parameters with or and .The respective edit window appears with Min and Max figures forthe parameter values Kpe, Tne, Tve, Tde and Xpf.− Change the value with or .− Single figures can be selected by pressing or .− Confirm your choice with the enter key .− By pressing (ESC) you are returned to the menu level without any change.F.2.1 Setting the correcting quantity limit− Access to the main menu level is obtained by pressing theenter key .− Selection and confirmation of Æ Setup Æ Control.The adjacent menu window appears.− Select and confirm Correction limit. with or and .The adjacent menu window appears. The minimum and maxi-mum possible values and the current value are displayed.− Change the value with or .− Single figures can be selected by pressing or .− Confirm your choice with the enter key .− By pressing (ESC) you are returned to the menu level without any change.ESC ּס OK - - -Correction limitMax: 500.0Min: 0.050.0ESC ּס OK StandbyExtern Pt100Kpe 0.5Tne 100Xpf 4.0Tv manual/auto autoTve 83(auto)Tde 83(auto)ESC ּס OK - - -KpeMax: 50Min: 0.10.5ESC ּס OK StandbyControlControl Variable ►Control parameter ►Control para. setsSetpoint offset ►Correction limit. 50.0.