ECO GOLD114 Appendix with settings 25/08/2011/ YACE0088Admissible data formats:-XXXX.XX -XXXX.X -XXXX. -XXXX XXXX.XX XXXX.X XXXX. XXXX-XXX.XX -XXX.X -XXX. -XXX XXX.XX XXX.X XXX. XXX-XX.XX -XX.X -XX. -XX XX.XX XX.X XX. XX-X.XX -X.X -X. -X X.XX X.X X. X-.XX -.X .XX .XG.5.2 Interface read commands (data request from the thermostat)Command MeaningIN_PV_00 Interrogation of bath temperature (outflow temperature)IN_PV_01 Interrogation of controlled temperature(int./ext., Pt/ext., Analog/ ext. serial).IN_PV_03 Interrogation of external temperature TE (Pt100).IN_PV_04 Interrogation of external temperature TE (Analog Input).IN_PV_10 Interrogation of bath temperature in 0.001 °C.IN_PV_13 Interrogation of external temperature TE (Pt100) in 0.0001 °C.IN_SP_00 Interrogation of temperature set value.IN_SP_01 Interrogation of pump power level.IN_SP_02 Interrogation of cooling mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).IN_SP_03 Interrogation of overtemperature switch-off point.IN_SP_04 Interrogation of outflow temperature limit TiH.IN_SP_05 Interrogation of outflow temperature limit TiL.IN_PAR_00 Interrogation of the control parameter Xp.IN_PAR_01 Interrogation of the control parameter Tn (181 = OFF).IN_PAR_02 Interrogation of the control parameter Tv.IN_PAR_03 Interrogation of the control parameter Td.IN_PAR_04 Interrogation of the control parameter KpE.IN_PAR_05 Interrogation of the control parameter TnE (response: XXXX; 9001 = OFF).IN_PAR_06 Interrogation of the control parameter TvE (response: XXXX; 5 = OFF).IN_PAR_07 Interrogation of the control parameter TdE (response: XXXX.X).IN_PAR_09 Interrogation of the max. correcting quantity limit.IN_PAR_10 Interrogation of the control parameter XpF.IN_PAR_14 Interrogation of setpoint offset.IN_PAR_15 Interrogation of the control parameter PropE.IN_DI_01 Status of Contact Input 1: 0 = open/ 1 = closed.IN_DI_02 Status of Contact Input 2: 0 = open/ 1 = closed.IN_DI_03 Status of Contact Input 3: 0 = open/ 1 = closed.IN_DO_01 Status of Contact Output 1:0 = NO contact open/ 1 = NO contact closed.IN_DO_02 Status of Contact Output 2:0 = NO contact open/ 1 = NO contact closed.IN_DO_03 Status of Contact Output 3:0 = NO contact open/ 1 = NO contact closed.IN_MODE_00 Keypad: 0 = released / 1 = locked.IN_MODE_01 Control: 0 = int./ 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. Serial.