CompressorA hermetically sealed compressor is used in the refrigerationunit. The compressor is equipped with overload protectionwhich trips on the compressor temperature and compressorcurrent consumption. CondenserDepending on the device type, an air-cooled or water-cooledcondenser is used in the refrigeration unit. In air-cooled con-densers, the condensation heat is discharged to the environ-ment. The fresh air is sucked in through the front of the deviceusing a fan, heated and discharged on the rear of the device.In water-cooled condensers, the condensation heat is dis-charged via cooling water. EvaporatorIn the internal bath, heat is discharged using a pipe coil evapo-rator.Technical information for the refrigeration unit can befound inÄ on page 61.3.4.6 InterfacesNote the following: The equipment connected to the low voltage inputs and out-puts must have safe separation from dangerous voltagesaccording to DIN EN 61140 such as by double or reinforcedinsulation according to DIN EN 60730-1 or DIN 60950-1.Specific functions of the device such as the setpoint temperaturecan be controlled with the RS232 interface using a PC. Thus,custom programs for controlling the device can be developed.Further information about the connection and configura-tion can be found inÄ Chapter 6.7.1 ‘ConfiguringRS232 interface’ on page 44.Changeover contact which switches in the event of a device fault.For example, faults can be registered on a system.Using the display, it can be set in which fault situationsa signal is output via the interface.RS232 interfaceAlarm outputDesign and functionMicrocool 21