Operation Manual MA 428100460 Rev.7 Ver.1 52 12/159. Cleaning and Disinfection RecommendationsIt is recommended that parts (device and accessories like headphones, earcushions) which come in direct contact with the patient be subjected tostandard cleaning and disinfecting procedure between patients. Pleasedisinfect the instrument and its accessories in dependence on the specificexisting infection potential.Recommendations for cleaning and disinfection of MAICO device presented inthis document are not intended to replace or contradict policies in effect orprocedures required for infection control at the facility.If there is not a high infection potential, MAICO recommends:Before cleaning always switch off and disconnect the device from powersupply.For cleaning use a lightly dampened cloth with soap water solutionDisinfect the instrument and its accessories by wiping the surfaces with wetSani-Cloth® Active wipes or a comparable product and allow them to takeeffect for the duration related to the specific disinfection aim listed in productdata sheet of the disinfection product. Please follow also its instructions forcleaning. Wipe before and after each patient After contamination After infectious patientsDisinfect computer, keyboard, transport trolley etc. with Sani-Cloth® Activewipes: once a week after contamination when pollutedTo avoid damage of the device and its accessories, pleasemind the following: Do not autoclave or sterilize. Do not use the device in the presence of fluid thatcan come into contact with any of the electroniccomponents or wiring.Should the user suspect fluids have contacted the systemcomponents or accessories, the unit should not be useduntil deemed safe by a MAICO certified servicetechnician.CAUTIONCAUTION