11 TY-GC62-010A Rev.2(2) Wall mountWhen screwing it to the mounting surface,tightening torque should be 1 N m or less.(3) DIN rail mountHang the hook on the back of the main unit to the DIN rail first, and press thebottom to fit the hook. To remove it, insert a slotted screwdriver into the groove ofthe stopper at the bottom and pull it.(Installation) (Uninstallation)6. Wiring and Connection■ 6.1 WiringFollow the figures to perform wiring. Use M3 terminal screw. Always use a crimp terminal(JIS C2805 Type R 1.25-3, or equivalent) for wiring. In that case, tightening torque to theterminal screw should be 0.5 N m or less.CAUTION Be careful not to cause an excessive impact on the product suchas hitting or dropping during installation of piping. Secure a space for maintenance activities.-1>-2>Slotted screwdriverStopperDIN railUnusedAnalog output GroundingOUT1 OUT2UnusedOUT1 OUT2 Output COMPower