L00049 1.0 08/2015 12Figure 12C2.4 Closing the Chamber Door The door will automatically lock at the start of a high level disinfection cycle. If the door is not properly closed, screen message: Close Chamber Door will be displayed.C2.5 Disinfecting the ProbeThe next screen message: Is the Probe Clean and Dry?1. If the probe has been pre-cleaned and dried, press Yes using the soft key button. If Yes isselected, you will be prompted by the screen message: Press start to begin.2. If the probe has NOT been pre-cleaned and dried, press No. Remove the probe, clean and drythe probe and position again as directed in section C2.3.3. Once the probe is pre-cleaned and dried, select Yes and press the START button to initiate thecycle or Cancel to unlock the machine and remove the probe.4. The progress of the disinfection cycle is indicated on the LCD screen message: Disinfecting.5. The High Level Disinfection cycle will take 7 minutes to complete.WARNING: In the unlikely event mist is visibly escaping from the chamber, remain at adistance from the trophon until completion of the operating cycle and the mist is nolonger visible. Do not come into direct contact with the mist. Contact your customerservice representative (see Part D – Troubleshooting).C2.6 Removing the Probe1. When the cycle has been completed, the trophon will sound an audible alarm.NOTE: After completion of a successful high level disinfection cycle, the ultrasound probe andchamber may have surface temperatures up to 45oC/ 113oF and 60oC/ 140oF respectively. Care