L00049 1.0 08/2015 13should be taken not to touch the chamber. The probe will feel warm to touch and is safe forhandling and use with gloves on.2. Follow the on screen instructions:Screen message 1: Cycle Complete. Wear Gloves. Remove and Wipe Probe.Screen message 2: Attention: Wear Gloves and Wipe Probe. (This message indicates thatsome hydrogen peroxide may not have been broken down into its constituents, oxygen andwater and extra care should be taken when removing the probe).3. Wearing clean gloves, immediately remove the used Chemical Indicator from the trophon andverify the colour change against the chart on the Chemical Indicator carton. Record the resultusing the printer or logbook accessories if required.4. Discard used Chemical Indicator.5. Remove the probe carefully using minimal contact after the cycle is complete. Avoid touchingthe probe against the chamber’s hot surface. Avoid compromising the probe’s disinfectedsurface before use.NOTE: After HLD cycle completion, the trophon performs a rapid cooling cycle until the probe isremoved from the chamber to prevent overheating of the probe. If the probe is not removedimmediately this will increase the warm-up time required by the subsequent cycle. It is thereforerecommended to remove the probe as soon as possible after HLD cycle has been completed.6. Wipe the probe with an absorbent, single-use, dry cloth/ wipe. Visually inspect the probe andensure any disinfectant residue present is removed.7. If a pass was verified by the Chemical Indicator colour AND the trophon screen displayedcycle complete, the HLD has been successful – proceed to step 8 below. If one or both ofthese items did not occur, repeat steps from section C2.1.8. Discard gloves.9. Close the chamber door.10. The probe is now ready for use or storage.C2.7 Sleep Mode and Shutdown Procedures If the trophon is not used for 120 minutes or a probe has been left inside the trophon for anextended amount of time, it will automatically enter sleep mode in order to save power. Torestart the trophon from sleep mode press Restart. While the trophon is in sleep mode it will perform self-maintenance functions and will displaythe messages: Warming Up or System Refresh. Please do not switch the trophon off duringthese processes. System refresh during sleep will only occur for low use customers and does not impact thenumber of disinfection cycles that can be performed per cartridge. This process will typicallytake 13 minutes.