Chapter 1 Installing and Configuring the SCXI-1129© National Instruments Corporation 1-15 SCXI-1129 User ManualTo make connections in the 12-slot and 12-slot multichassis configuration,refer to Figure 1-6 while completing the following steps:Figure 1-6. 12-Slot to 12-Slot Multichassis Configuration Parts Locator Diagram1. Install the HVAB-backplane adapters and jumper blocks in the 12-slotchassis as described in steps 1 through 3 in the Single 12-Slot ChassisConfiguration section.2. Connect the HV8-HV8 cable from the first chassis, normally theconnector behind slot 1, to the HVAB connector (behind slot 12) of thenext chassis.3. Connect the SH9MD-9MD cable from the AUX OUT connector of thefirst chassis to the AUX IN connector (behind slot 5) of the nextchassis.4. Repeat steps 1 through 3 for each additional chassis.1 8-Slot HVAB-Backplane Adapter2 HV8-BAN4 Cable3 8-Position HVAB Plug4 2-Slot HVAB-Backplane Adapter5 1-Slot HVAB-Backplane Adapter6 NI 4060 for PCI7 SH9MD-9MD Cable8 12-Slot SCXI Chassis9 HV8-HV8 Cable10 12-Slot SCXI Chassis11 SH9MD-9MD CableAB0+AB0–AB2+AB2–6897101 32114 5