-27-7.2.2.RS-232C communication command•Typical communication command"cr"means carriage return.• Other communication commandOther than those above, a lot of communication commands are prepaired.Transmittingcommandfrom hostcomputer toFGPContent Returning commandfrom FGV ExplanationAAcr Tare AAcrABcr Cancel of data transmission ABcrACcr Switch to plus peak hold mode ACcrADcr Switch to standard measuring mode ADcrALcr Switch to minus peak hold mode ALcrAEcr Clear the plus/minus peak value tozero AEcrAFcr Switch the unit to kg AFcrAGcr Swithc the unit to N AGcrAHcr Swithc the unit to lb AHcrAKcr Swithc the unit to oz AKcrBAcr Transmission request of one measuringdata (measuring valus at present) BAcr NAoooooocroooooo :6-digit value including sign, decimal point and 4-digitnumberBBcr Request for continuous transmission ofmeasuring data (10 times/second) BBcr NAoooooocrBB1cr Request for continuous transmission ofmeasuring data (20 times/second) BB1cr NAoooooocrBB2cr Request for continuous transmission ofmeasuring data (50 times/second) BB2cr NAoooooocrBB3cr Request for continuous transmission ofmeasuring data (100 times/second) BB3cr NAoooooocrBCcr Transmission request of model BCcr NEoocroo : 2-digit number indicating modelOC : FGV - 500HXYOD : FGV - 1000HXYBDcr Transmission request of unit BDcr NHocr o : one-digit number indicating unit0 : N, 1 : kg, 2 : g, 3 : lb, 4 : ozBEcr Transmission request of plus peakvalue BEcr NBoooooocr oooooo :6-digit value including sign, decimal point and 4-digitnumberBFcr Transmission request of minus peakvalue BFcr NCoooooocrIn the communication with host computer, whenFGV detect communication error, transmit the errorcommand.OBcr Command format error (mistake command)OFcr Flaming errorOHcr Overrun error