-29-Available to output one measuring data to the device (printer) provided Mitutoyo Digimatic communication.When the peak or hold mode are ON, just one hold data is outputted.*In function setting: f06, need to set "dig". And then, start Digimatic communication.If setting is "Ser", you cannot communicate with Digimatic.*Please make the connector to meet the device.Signal connectorFGV-o :Cable for the signal connector(option)8. Signal Connector8.1. Pin assignment8.2. Mitutoyo Digimatic OutputPin number Signal name1 REQ signalDigimatic2 READY signal3 CLK signal4 DATA signal5 Comparator upper limit outputComparator6 Comparator OK output7 Comparator lower limit output8 ZERO signal inputInput signal9 HOLD signal input10 PEAK signal input11 Common for comparator output12 GNDThe device (printer) providedfor Mitutoyo Digimatic*Please purchase the optionalspecial cable when using thesignal connector.Doing comparator output.• I/F circuit diagram8.3 Comparator output• Comparator outputWhen a comparator max output is ON, the photo coupler of PC3 is turned on and current flows.When a comparator OK output is ON, the photo coupler of PC4 is turned on and current flows.When a comparator min output is ON, the photo coupler of PC5 is turned on and current flows.Please refer to "5.5. Comparator" for the details (the setting method of comparator upperlimit and a lower limit, the judgment conditions of a comparator) of a comparator function.Comparator max outputComparator OK outputComparator Min outputCommon