BioMonitor SmartLeadP852 Rev B (April 2019) 11If a SmartLead is connected to a sensor when the sensors are detected in the MR3 hardwaresetup (Section 6 -> MR3 Configuration) it will override the EMG functionality of that sensor.Therefore, if the SmartLead is moved to a new sensor, Step 1-5 of Section 6 (MR3 Configuration)must be recompleted.To check if the SmartLeads are currently connected to the proper Ultium sensor, the Find MySensor feature may be useful. Click the topmost star (shown in the above figure). If all lights blink(white color), then they are properly connected. If one of the sensors blinks (red color), theSmartLead that is connected to this sensor should be connected to another sensor. It isrecommended to recomplete Steps 1-5 of Section 6 above if this does occur.7.1 Normal Appearance of SignalsThe sensor’s STATUS LED provides a means of communicating its operational state. In the idlestate, the STATUS LED will flash blue at a low, once per second rate. When the sensor is activelymeasuring a signal, the STATUS LED will flash recognizably faster (green).7.2 Quick Testing:Place the sensor on the subject as instructed in Appendix A and visually inspect the signal.Depending upon which attachment configuration is used the BioMonitor signals should resemblethe image below.7 Pre-Use Check-Out