Comprehensive Manual78© 2018 Nortek AS3.3.7.1 Midlife ElectronicsMidlife electronics change the power levels with a simple command in the terminal emulator:Voltage level Command Max currentRAW OPWRBA 500 mA+5V OPWR5V 250 mA+12V OPWR12 100 mA3.3.7.2 Pre-midlife ElectronicsYou can change the output voltage level that is supplied to external sensors. This is typically neededif an external analog sensor (e.g. OBS) operates in a specific voltage range and the raw batterypower does not fall within this range. The default output voltage level is RAW (battery or externaladapter). This can be changed to +5V or to +12V by changing the position of a 0-W resistor. Thethree resistor positions share one pad so the modification should be straightforward.Voltage level Resistor position Max currentRAW R270 (default) 500 mA+5V R268 200 mA+12V R269 50 mA