NOUVAG AG MD 10No. 31820 22/07 5Insure that „pressure“ tube runs to irrigation needle, and „suction“ tube runs tosolution bottle!10) Insert tube needle and hang cooling flask on hanger rod.11) Open tube clamp.12) Open air valve of the drip chamber.13) Connect the mains cable into the electrical outlet.Check for correct operation voltage and verify mains voltage output!2304.2 Tubing assembly-Internal irrigation1) Insert internal irrigation needle (1) into the end of thetubing.2) Attach clip (2 & 3) to irrigation tubing.3) Insert irrigation needle into contra angle.4) Attach clip to contra angle and to motor cable.5) When necessary, attach additional clips to motor cable.1234.3 Tubing assembly–External irrigation1) Attach clip (1 & 2) to irrigation tubing.2) Connect piece of tubing to irrigation needle (3).3) Attach clip to contra angle and to motor cable.4) When necessary, attach additional clips to motorcable.3 124.4 Tubing assembly-combined1) Attach Y-connector (1) to the end of the tubing.2) Attach two 16cm pieces of tubing (2) to Y-connector.3) For internal irrigation, connect first tubing piece toirrigation needle (3).4) Attach clips (4 & 5) to irrigation tubing.5) Attach irrigation tubing with clip to contra angle and tomotor cable.6) Insert irrigation needle into contra angle.7) Connect second tubing piece with external irrigation.8) When necessary, attach additional clips to motor cable.45123