MD 10 NOUVAG AG6 No. 31820 22/075 Operation5.1 Starting the MD 10Press green “ON/OFF“ switch.5.2 KeyboardA) Display: Indication of the revolutions (speed), torque or pump speed.B) Multiplying key “1:5” (for contra angle 1:5):Speed: 2,500 – 200,000 rpm Torque: 1, 2 NcmC) Multiplying key “1:1” (for contra angle 1:1):Speed: 500 – 40,000 rpm Torque: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 NcmD) Reduction “16:1” (for contra angle 16:1):Speed: 30 – 2,500 rpm Torque: 5, 7, 10, 12, 15, 17, 20, 25, 30, 35 NcmE) Reduction “32:1” (for contra angle 32:1):Speed: 15 – 1,250 rpm Torque: 10, 15, 20, 25, 30, 35, 40, 45, 50, 55 NcmF) Fixed key “SAW” (for Nouvag saws):Speed: 14,000 rpm Torque: 10 NcmG) Fixed key “MUCO” (for Nouvag Mucotome):Speed: 8,000 rpm Torque: 10 NcmH) Key “ATC SET”:Press key “ATC SET” to adjust the torque limiter AL or AS.I) Key “PUMP SET”:Press key “PUMP SET” to adjust the pump speed.J) Taste “DOWN”:To decrease the speed, torque and pump.K) Taste “UP”:To increase the speed, torque and pump.L) Light “MOTOR ON”:The green light indicates that the electronicmotor is switched on.M) Light “REVERSE MODE”:The green light indicates that the electronicmotor operates in reverse and a warning tone will sound.5.3 Setting the revolutions-1) Press the appropriate reduction or multiplying key (for example key “32:1” for contra angle 32:1). The green light on key B) to I) illuminates, when the key is activated.16:1 32:1