6.5. ìSuspendî modeIf the measurement and the corresponding palliativemeasurement fail four times consecutively the DiasysIntegra Access goes into ìsuspend modeî.The measurements will be carried out according to theprogrammed cycle, but without the palliativemeasurements. Also, the inflation level will be 150mmHg.As soon as a valid measurement is recorded, the unit comesout of ìsuspendî mode and the recording continues asinitially programmed.6.6. Measurements which are not memorisedBy pressing the two keys simultaneously, the physician cantrigger, at any time, a measurement which will not bememorised (measurements are displayed only) withoutaffecting the ongoing program.6.7. Display of measurementsAfter each measurement the unit displays successively:Stop RecOn/ValidSelectPress Valid and Select simultaneouslyto trigger a measurement which willnot be memorised.30 ï Diasys Integra Access - Manual - GB - Rev 2Measurement modeSystolic pressureDiastolyc pressureHeart rateStandard function: