EN-213000 Series Indicators3.5.3 Auto-Zero TrackingSet the automatic zero tracking functionality.OFF = disabled.0.5 d = the display will maintain zero until a drift of 0.5 divisions per second has beenexceeded.1 d = the display will maintain zero until a drift of 1 division per second has beenexceeded.3 d = the display will maintain zero until a drift of 3 divisions per second has beenexceeded.NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d and 3d. The setting islocked when the hardware lock switch is set to the ON position.3.5.4 BacklightSet the display backlight functionality.OFF = always off.ON = always on.AUtO = turns on when a button is pressed or the displayed weight changes.turns off after 5 seconds of no activity.3.5.5 Auto Off TimerSet the automatic shut off functionality.OFF = disabledSEt 1 = powers off after 1 minute of no activity.SEt 2 = powers off after 2 minutes of no activity.SEt 5 = powers off after 5 minutes of no activity.3.5.6 End ReadoutAdvance to the next menu.3.5.2 FilterSet the amount of signal filtering.LO = less stability, faster stabilization time (<1 sec.)MEd = normal stability, stabilization time (<2 sec.)HI = greater stability, slower stabilization time (<3 sec.)www. GlobalTestSupply .comFind Quality Products Online at: sales@GlobalTestSupply.com