EN-26 Adventurer ProResetSet On to lock the Reset Menu.Lock SetSet On to lock all submenu lock settings. Whenset Off, the lock settings for all menus areaccessible. To turn Lock Set off, refer to section3.7.13.End LockoutPress the Yes button to advance to the Endmenu. Press the No button to return to the Calmenu.3.7.12 EndThe End menu is used to exit the menus andreturn to previous application3.8 Legal for Trade (LFT)Specific Adventurer Pro models have beendesigned to comply with OIML, EEC, NTEP andMeasurement Canada weights and measuresregulations. These approvals are pending; socontact Ohaus for further details aboutavailability.When the LEGAL TRADE menu item setting isset to ON, the following conditions apply: The CALIBRATION menu is hidden. The LEGAL TRADE menu item is hidden. The AUTO ZERO menu item setting is setto 0.5d and locked. The OutPUt WHENSTABLE menu itemsetting is set to ON and locked. The CONTINUOUS setting in the AUTOPRINT menu item is hidden.When used in conjunction with the Lockoutmenu, the balance may be sealed to preventor detect unauthorized changes to the balancesettings. For legal for trade applications, thebalance must be sealed to prevent access tothe metrological parameters.To seal the balance, cover the Lock switchaccess hole at the rear of the balance. A paperseal, and a wire seal may be installed asshown.To regain access to the locked balance settings,break the seal and press the Lock switch whileOHAUS is displayed during power up.3.9 Sealing Access to theBalance SettingsSealing the BalanceSEALWire Seal Paper SealRear of the Balance4 Adventurer Pro Text pages 15 to 40.P65 4/10/2003, 6:00 AM26