CHAPTER 1 GETTING STARTEDOhaus Corporation www.ohaus.com 1-4 EB Series Scales Service Manual1.6 SCALE SETTINGSThe EB scale contains four user-selectable scale settings (Setup mode). These are as follows:Scale Increment, Backlight, Zero Tracking Range, and Filtering. A description of each setting isexplained below.1. To enter the setup mode, press and then release COUNT and M+ at the same time toenter into the user-selectable scale settings.2. In the Setup mode, press ▲ to step through available settings.3. Press Enter to accept the displayed setting and proceed to the next parameter.4. Press Exit to proceed to the next parameter without saving any changes.Table 1-3. MENU STRUCTUREinC 01 EL AU AC 2d Fil 1AUONOFF0.5d1d2d4d0123(Levels)(d=scale division)(mode-dependentsettings)inC 01EL AUAC 2dFiL 1Scale increment - Sets the displayed scale increment (also known as readability orgraduation). Selectable setup values will be model dependent and are equivalentto within 30,000 to 3,000 scale divisions.Backlight - Sets the activation mode of the backlight.The following settings are available: AU (Auto-on with items > 9d placed on thePlatform or any button pressed; turns off after 5 seconds of inactivity), On, oFF.Zero Display Range - Sets the range in which the Zero reading is maintained.The following settings are available: 0.5d, 1d, 2d, 4d (d= scale division).Filtering - Sets the level at which the Stable indication turns on; the higher thesetting, the faster the stabilization time.The following settings are available: 0, 1, 2, 3 (levels).