EC Series EN-32 INSTALLATION2.1 UnpackingUnpack and verify that the following components have been included:; EC Scale; Weighing Platform (plastic base with stainless steel pan); Instruction Manual; AC AdapterSave the packaging material. This packaging ensures the best possible protectionfor the storage or transport of the product.2.2 Installing ComponentsPosition the weighing platform pins into the through-holes on the top housing, thenset the platform securely into place. Secure the in-use cover with double-sided tapeas needed.IMPORTANT:; BEFORE USING THE SCALE, REMOVE THE SHIPPING PROTECTION SCREWlocated underneath the scale. This screw ensures protection of the load cellduring transport, but will need to be removed for the scale to operate properly.; It is recommended that the protective screw be re-installed if transporting thescale. Do not over-tighten the protective screw as it may damage the load cell.To re-install, turn the scale on and place a 500g weight on the pan, thenslowly screw in the protection screw until the display shows a value between250g to 400g.2.3 Selecting the LocationOperate the scale on a firm, level surface. Avoid locations with rapid temperaturechanges, excessive dust, moisture, air currents, vibrations, electromagnetic fields,heat or direct sunlight.2.4 Leveling the ScaleAdjust the leveling feet until the bubble is centeredin the circle of the level indicator(located on the front panel).NOTE: Ensure that the scale is level each time its location is changed.3 EC-textEN.pmd 8/9/2005, 1:41 PM3