8 GETTING STARTED1.4 SERVICE FACILITIESTo service a centrifuge, the service area should meet the following requirements:• Area must be clean and air must not contain excessive dust particles.• Work surface must be stable and level.• Use an approved Electro-Static Device.1.5 TOOLS AND TEST EQUIPMENT REQUIRED1.5.1 Standard Tools and Test Equipment1. Flat and Philip screws drivers, tweezers, adjustable open wrenches and socket set (withdrive socket sets from 4mm-14mm, Ratchet and drive extension bar.)1.6 Transporting and Packing Centrifuge FC57141.6.1 Transport• Please transport the device in the original packaging.• Use a transport aid for transporting over longer distances.1.6.2 PackingPack the centrifuge in reverse order using the equipment original packing.1.7 TECHNICAL DATA1.7.1 Acceleration times FC5714 (120 V / 230 V) in min/secRotor-NumberLevel 0Level 1Level 2Level 3Level 4Level 5Level 6Level 7Level 8Level 930314823 1:40 0:52 0:35 0:26 0:21 0:18 0:16 0:15 0:13 0:1130314834 1:42 0:50 0:34 0:26 0:21 0:18 0:15 0:14 0:14 0:1430314830 1:41 0:51 0:34 0:26 0:21 0:18 0:15 0:14 0:14 0:1430314822 1:52 0:57 0:39 0:29 0:24 0:20 0:18 0:16 0:14 0:1430314824 3:48 1:55 1:19 1:01 0:47 0:40 0:35 0:30 0:28 0:2530314836 4:13 2:01 1:21 1:02 0:50 0:42 0:36 0:32 0:29 0:2630314826 6:21 3:09 2:08 1:37 1:16 1:05 0:57 0:49 0:44 0:4030314832 2:36 1:16 0:50 0:39 0:32 0:27 0:24 0:21 0:19 0:19