Ranger™ 4000 Series EN 16Reset the Read menu to factory defaults.NO = not resetYES = resetStable Range [Stable]Set the amount the reading can vary while the stability symbol remains on.0.5d = 0.5 scale division1d = 1 scale division2d = 2 scale division5d = 5 scale divisionFilter [FILtEr]Set the amount of signal filtering.LOW = less stability, faster stabilization timeMED = normal stability, stabilization timeHI = greater stability, slower stabilization timeAZT [AZt]Set the automatic zero tracking functionality.OFF = disabled0.5d = the display will maintain zero until a change of 0.5 divisions per second hasbeen exceeded.1d = the display will maintain zero until a change of 1 divisions per second has beenexceeded.3d = the display will maintain zero until a change of 3 divisions per second has beenexceeded.Light [LIGHt]LCD models:Sets backlight functionality.OFF = always offON = always onAUTO = turns on when a button is pressed or the displayed weight changes.LED models:Set the light intensity.LO = low light intensityMED = medium light intensityHI = high light intensitySleep [SLEEP]Set the display shut off time.OFF = disabled (display will stay on until power off)ON = display will turn off in one minute in the absence of key or weight actionAuto off [A.OFF]Set the automatic shut off functionality.OFF = disabled1 = powers off after 1 minute of no activity5 = powers off after 5 minute of no activity10 = powers off after 10 minute of no activity