Maintaining the unit4-1 6600-0361-000 08/10/02 4-1Repair PolicyWarranty repair and service must be performed by an OhmedaMedical Service Representative or at the Ohmeda MedicalService and Distribution Center. To contact an Ohmeda MedicalService Representative, call the your Ohmeda Medical ServiceOffice listed on the back cover.Do not use malfunctioning equipment. Make all necessaryrepairs or have the equipment repaired by an Ohmeda MedicalService Representative. Parts listed in the service manual forthis product may be repaired or replaced by a competent,trained person who has experience in repairing devices of thisnature. After repair, test the equipment to ascertain that itcomplies with the published specifications.CAUTION w Detailed information for more extensiverepairs is included in the service manual solelyfor the convenience of users having properknowledge, tools and test equipment, and forservice representatives trained by OhmedaMedical.Maintenance scheduleThe unit should be maintained in accordance with the proce-dures detailed in the Service Manual. Service maintenance mustbe performed by a technically competent individual.Operator maintenanceThis schedule lists the minimum maintenance frequencies.Always follow hospital and local regulations for required mainte-nance frequencies.Weekly or After Each PatientClean the Giraffe Spot PT Lite Phototherapy System. Disinfectthe unit if required or after use with infectious patients.QuarterlyInspect the air filter and clean or replace as required.Note: This is the minimum inspection frequency. The filter mustbe cleaned or replaced whenever it appears dirty.