6.5 Reprocessing the endoscopic CO2 regulation unit and accessories59UCR INSTRUCTION MANUALCh.6 Reprocessing workflow for the endoscopic CO2 regulation unitand accessoriesThis chapter describes the workflow for reprocessing the endoscopic CO2 regulation unit andaccessories.6.5 Reprocessing the endoscopic CO 2 regulationunit and accessories1 Confirm the appearance of the endoscopic CO2 regulation unit and accessories.2 If patient debris directly or indirectly adheres to this instrument and accessories, referto “ Disinfection” on page 60. Otherwise, be sure to follow the next step.3 If disinfection is necessary according to national, local, or facility standards, refer to“ Disinfection” on page 60. Otherwise, refer to “ Cleaning” on page 61.3. Disinfection is necessary according tonational, local, or facility standards2. Patient debris directly or indirectly adheres to this instrumentCleaning(→”Cleaning”)Disinfection(→”Disinfection”)YesYes NoNo1. Confirm the appearance of the endoscopic CO 2 regulation unit