Transformer6 - 15Test settingsFigure 10:RWinding test cardwith test resultsNavigate to the parameter fields, and enter the values according to your testrequirements:Range: output rangeNote: The 400A DC output range provides maximum current 100 A.Itest: nominal test currentR min: calculated minimum winding resistance value (display only).Depends on the value of the nominal test current and themeasuring range:– 400A DC: Rmin = 0.2mV / Itest– 6A DC: Rmin = 0.2mV / Itest– V DC (2 wire): Rmin = 0.2ΩR max: calculated maximum winding resistance value (display only).Depends on the value of the nominal test current and themeasuring range:– 400A DC: Rmax = 5V / Itest.– 6A DC: Rmax = 10V / Itest.– V DC (2 wire): Rmax = 20kΩ