219. WEIGHINGBefore start of weighing process or in case of essential change of ambientconditions at a workstation (e.g. ambient temperature change at a workstation morethan 3 o C) the balance requires adjusting. the procedure of balance adjustment isdescribed further in this user manual.• Before start of weighing procedure, it is recommended to load the balance’sweighing pan a few times with mass close to balance max capacity,• Check if unloaded balance indicates “precise zero”- and whethermeasurement is stable - , if not press ESC/TARE key;• Press Units key to set a measuring unit:[g], [mg], [ct], and also if enabled in factory menu [kg], [lb], [oz], [ozt], [dwt],[t], [mom], [gn],• place weighed object on balance’s weighing pan and read result only onstabilization of measurement result,• mass indication of a load placed on balance’s weighing pan can be tarred formultiple times by pressingESC/TARE key ((pay attention not to exceedmaximal capacity of a balance by applying multiple tare function).During times between carrying out the following measurement series do not unplugthe balance from mains. It is recommended to switch off balance’s display bypressing ON/OFF key. On repeated pressing of theON/OFF key the balance isready for operation and carrying out the following measurements.