7826.4.2. Continuous transmissionResults are transmitted in basic units. Transmission can be disabled/enabledusing RS232 command:C1 CR LF – Switch on continuous transmission in basic unitC0 CR LF – Switch off continuous transmission in basic unitor using the parameter: settings/ RS-232/Aut. printout/continuousThe frame length is constant (21 characters).Frame format:[SI] [space] [stability] [space] [sign] [mass] [space] [unit] [CR LF]Stability character – [space] if stable[?] if not stable[^] if an indication over the range[v] if an indication below the rangesign – [space] for positive values or [-] for negative valuesmass – 9 characters justified to the rightunit – 3 characters justified to the left26.4.3. Continuous transmission „with pauses”Continuous transmission „with pauses” allows users to determine the timeinterval between the subsequent frames in the parameter interval.His type of printouts can be enabled only in parameters: Settings/RS-232/ Aut.printouts/with interv.The frame format is the same as in basic simple transmission.The interval is set in: Setting/RS-232/Interval26.4.4. Stable result transmissionThe scale automatically generates after each stable result separated by astable result below the Min. mass that can be set in parameters.The intended use of this printout is for slip printers so it doesn’t have a startingcharacters and the length can vary from 18 to 20 characters.