3155 User ManualPerformance Verification 5-73. Change the Model 3155 FREQ setting to 50.00 MHz(sine wave output to 100MHz). Verify that peak-to-peakof the displayed sinewave is greater than 5 divisions.Trig, Gate andBurst Characteris-ticsTriggered - Each transition at the front panel TRIG IN connec-tor stimulates the Model 3155 to generate one complete outputwaveform.Gated - External signal at the TRIG IN connector enables theModel 3155 output. The last cycle of the output waveform isalways completed.Burst - Each transition at the front panel TRIG IN connectorstimulates the Model 3155 to generate a burst of pre-selectednumber of cycles.Equipment: Pulse/function generator, oscilloscope.Trigger1. Configure the Model 3155 as follows:Function: Sine waveFrequency: 1.000 MHzOperating Mode: TriggeredRemote Commands: :RES:OUTP ON:FREQ 1e6:INIT:CONT OFF2. Set the external pulse/function generator fre-quency to 10 kHz and connect its SYNC output tothe trigger input of the 3155. Set the oscilloscopeand verify on the oscilloscope that the Model 3155outputs a triggered signal. Leave the externalpulse generator connected to the Model 3155 forthe next test.Gate1. Configure the Model 3155 as follows:Function: Sine waveFrequency: 1.000 kHzOperating Mode: GatedRemote Commands: :RES:OUTP ON:FREQ 1e3:INIT:CONT OFF:GATE:STAT ON2. Set the oscilloscope and verify that the Model