MultiRAE User’s Guide5416. Once you have made your selections in ProRAE Studio II, you must upload the changes to theinstrument. Click the icon labeled “Upload all settings to the instrument.”17. A confirmation screen is shown. Click “Yes” to perform the upload, or “No” to abort.Uploading takes a few seconds, and this progress bar is shown. You can abort the upload by clicking“Cancel.”18. Exit ProRAE Studio II.19. Press [Y/+] on the MultiRAE to exit Communication Mode.9.2 Deactivating Policy Enforcement9.2.1 AutoRAE 2 CradleTo deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the procedure for changingsettings. See page 49 for details.9.2.2 MultiRAE Desktop Cradle Or Travel ChargerIf the MultiRAE screen displays the message that it must be bump tested or calibrated, and if the option tobypass bump testing or calibration is not available, you should shut off the instrument and follow theprocedure outlined here if you want to change the Policy Enforcement settings:1. Use a USB cable to connect the MultiRAE in its Travel Charger or Desktop Cradle to a computerrunning ProRAE Studio II.2. Enter Diagnostic Mode on the MultiRAE (with the instrument turned off, press and hold [Y/+] and[MODE] until it starts up.3. After startup, enter the password when prompted (default is “0000”) and press [MODE].4. Press [N/-] repeatedly until you see the “Enter Communications Mode?” screen.5. Press [Y/+] to enter Communications Mode.6. Start ProRAE Studio II.7. Select “Administrator.”