Wall Mounting InstructionsWARNING: IT IS IMPORTANT TO SAFELY SECURE THE EQUIPMENT. UNSECURED EQUIPMENT COULDPOSSIBLY BECOME DISLODGED AND FALL, CAUSING INJURY TO EITHER THE PATIENT OR EXAMINER.WARNING: CARE MUST BE TAKEN TO ARRANGE THE CABLES FOR THE ACCESSORIES SUCH THATTHEY DO NOT PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT.WARNING: POSITION THE CLEARCHART 4 ON THE WALL SO THAT IT IS NOT DIFFICULT TO OPERATETHE DISCONNECTION DEVICE (PLUG).Note: Make sure the position you choose to mount your ClearChart 4 is within the reach of a poweroutlet.Direct-Throw: Your ClearChart 4 must be positioned at patient eye level directly in front of the patient.The minimum test distance is 6 feet (1.83 meters) with a maximum test distance of 24.6 feet (7.5meters). Refer to the Direct Throw illustration in this section.Mirror Arrangement: A first surface mirror is a useful space-saving device to increase the patienttesting distance when the room does not permit a direct-throw arrangement. Typically, your ClearChart4 will be positioned higher than patient eye level so that the examiner will not interfere with the patient’sview of the ClearChart 4. Arrange the ClearChart 4 and mirror so that the ClearChart 4 can be seen bythe patient through the mirror.1. Find an appropriate spot on the wall that will support the ClearChart 4, which will hang from theVESA compliant Mounting Bracket. Ensure that the refraction distance meets the requirementsfor either Direct Throw or Mirror Arrangement.2. Level the supplied Wall Mount Bracket and mark the holes on the wall. The mount should beinstalled in solid wood stud, brick or concrete wall.3. Hang the ClearChart 4 by lining-up the holes on the back of the device with the four holes on themounting bracket. Secure the ClearChart 4 to the mounting bracket with the 4 screws provided.13795-101 Rev. F9Installation, Features, & Functions (continued)