User's Guide _______________________________________________________________________40 ___________________________________________________________________ M211060EN-GOutputting the results continues in intervals issued with the commandINTV. You can stop the output by entering the S command.Stop Measurement OutputUse the S command or press the Esc key to stop the continuousmeasurement output.SOutput the Measurement Message OnceUse the SEND command to output the measurement values once. If theprobe is in POLL mode and the line is not open for commands, specifythe address of the probe to receive the measurement message.SEND [aaa]whereaaa = Address of the probe, range 0 ... 255. Set with the ADDRcommand.Example (probe in STOP mode, no address needed):sendT= 22.7 'C RH= 20.0 %RH Td= -1.5 'CExample (probe in POLL mode, with address 10):send 10T= 22.8 'C RH= 20.1 %RH Td= -1.3 'CConfiguring Serial Line OperationSet Serial Line SettingsUse the SERI command to show or set the serial line settings. The newsettings will be taken into use when the probe is reset or powered up.SERI [b p d s]whereb = baud rate(300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600)p = parity (n = none, e = even, o = odd)d = data bits (7 or 8)s = stop bits (1 or 2)