User's Guide _______________________________________________________________________48 ___________________________________________________________________ M211060EN-GOther CommandsSet Analog Output ModeUse the AMODE command to show or set the operation mode of theanalog output. This command is not available on the HMP110D.AMODE [ch1] [ch2]wherech1 = Analog output mode for channel 1, range 0 ...3.The options are:0 (0 ... 1 V)1 (0 ... 2.5 V)2 (0 ... 5 V)3 (1 ... 5 V)ch2 = Analog output mode for channel 2, range 0 ...3.The options are the same as for channel 1.Example (show current output modes):amodeCh1 output : 0 ... 1 VCh2 output : 0 ... 1 VExample (set channel 1 to 0 ... 1 V and channel 2 to 0 ... 5 V):amode 0 2Ch1 output : 0 ... 1 VCh2 output : 0 ... 5 V