73 MountingVEGAFLEX 82 • Four-wire 4 … 20 mA/HART47593-EN-1907013 Mounting3.1 General instructions for use of the instrumentProtect your instrument against moisture ingress through the followingmeasures:• Use a suitable connection cable (see chapter "Connecting topower supply")• Tighten the cable gland or plug connector• Lead the connection cable downward in front of the cable entry orplug connectorThis applies mainly to outdoor installations, in areas where highhumidity is expected (e.g. through cleaning processes) and on cooledor heated vessels.Note:Make sure that the degree of contamination specified in chapter"Technical data" meets the existing ambient conditions.Note:Make sure that during installation or maintenance no moisture or dirtcan get inside the instrument.To maintain the housing protection, make sure that the housing lid isclosed during operation and locked, if necessary.3.2 Mounting instructionsMount VEGAFLEX 82 in such a way that the distance to vessel instal-lations or to the vessel wall is at least 300 mm (12 in). In non-metallicvessels, the distance to the vessel wall should be at least 500 mm(19.7 in).During operation, the probe must not touch any installations or thevessel wall. If necessary, fasten the probe end.In vessels with conical bottom it can be advantageous to mount thesensor in the center of the vessel, as measurement is then possiblenearly down to the lowest point of the bottom. Keep in mind thatmeasurement all the way down to the tip of the probe may not be pos-sible. The exact value of the min. distance (lower dead zone) is statedin chapter "Technical data" of the operating instructions.Fig. 2: Vessel with conical bottomProtection against mois-tureInstallation position